    題名: 誤差、線性化和取樣如何影響核磁共振訊號的初始值和時間常數
    How Errors, Linearization, and Sampling Affect Initial Values and Time Constants of NMR Signals
    作者: 林宜興
    Yixing, Lin
    貢獻者: 蔡尚岳
    Tsai, Shang-Yueh
    林宜興
    Lin Yixing
    關鍵詞: 核磁共振譜學
    代謝物濃度
    代謝物橫向鬆弛時間
    回訊時間
    Nuclear Magnetic Resonance Spectroscopy
    Metabolite Concentration
    Metabolite T2
    Echo Time
    日期: 2025
    上傳時間: 2025-05-02 15:05:43 (UTC+8)
    摘要: 研究目的
    從多個回訊時間所測得的訊號大小來擬合出代謝物濃度可以避免掉只從單一個回訊時間與單一個固定的時間常數去回推代謝物濃度的不準確之處,但是用擬合的方式也有三個面向要考慮:在回訊時間所測得的訊號大小會有各種誤差、不同的擬合方法可能會計算出不同的初始濃度與時間常數、也許不需要用到很多個回訊時間就可以有不錯的準確度。
    研究方法
    我們用程式模擬的方式來探討誤差、線性化和取樣如何影響代謝物的初始濃度和時間常數。我們針對時間常數是五十毫秒到四百毫秒的代謝物,施加百分之零點五到百分之十的測量誤差,並觀察訊號大小在兩組不同的回訊時間下有何變化。
    研究結果
    關於線性化,應透過非線性擬合而不是線性擬合來計算代謝物的初始濃度和時間常數,因為當存在測量誤差時,非線性擬合計算出來的初始濃度誤差較小,時間常數誤差也較小。關於誤差,初始濃度的誤差約為測量誤差的一半,時間常數的誤差約為測量誤差的兩倍。關於取樣,若將回訊時間從八個減少到四個,則初始濃度的誤差增加不到測量誤差的一半,時間常數的誤差增加不到測量誤差的兩倍。
    Purposes
    Fitting from multiple echo times to obtain the concentration of a metabolite circumvents the inaccuracy of extrapolation from just one echo time and from a rigid time constant to obtain the concentration, but three considerations regarding fitting arise: Echo signals picked up by the induction coil suffer various sources of errors, different fitting methods may compute different initial concentrations or time constants, and the number of echo times needed may possibly be reduced for efficiency.
    Methods
    We run simulations to see how errors, linearization, and sampling affect the initial concentration and the time constant of a metabolite having a T2 between 50 ms and 400 ms if there is a fixed measurement error between 0.5 % and 10 % from the two sets of echo times: 30, 50, 74, 101, 135, 179, 241, 350 ms and 1, 2, 5, 13, 30, 72, 170, 400 ms.
    Results
    Regarding linearization, the initial concentration and the time constant of a metabolite should be computed by nonlinear fitting instead of linear fitting because nonlinear fitting produces a smaller error of initial concentration and a smaller error of time constant when there is a measurement error. Regarding errors, the error of initial concentration is approximately one half of the measurement error and the error of time constant is approximately twice the measurement error. Regarding sampling, reducing the number of echo times from the set {1, 2, 5, 13, 30, 72, 170, 400} to the set {1, 7, 54, 400} increases the error of initial concentration by less than one half of the measurement error and increases the error of time constant by less than two times the measurement error.
    參考文獻: [1] Watson, P. E., Watson, I. D., & Batt, R. D. (1980). Total body water volumes for adult males and females estimated from simple anthropometric measurements. The American Journal of Clinical Nutrition, 33(1), 27–39.
    [2] Cottingham, W. N., & Greenwood, D. A. (2001). An Introduction to Nuclear Physics (2nd ed.). Cambridge University Press.
    [3] Griffiths, D. J. (2013). Introduction to Electrodynamics (4th ed.). Pearson.
    [4] Serway, R. A., & Jewett, J. W. (2014). Physics for Scientists and Engineers with Modern Physics (9th ed.). Brooks/Cole Cengage Learning.
    [5] Levitt, M. H. (2001). Spin Dynamics: Basics of Nuclear Magnetic Resonance. Wiley.
    [6] Rabi, I. I., Zacharias, J. R., Millman, S., & Kusch, P. (1938). A new method of measuring nuclear magnetic moment. Phys. Rev. 53, 318.
    [7] Hoult, D. I., Bhakar, B. (1997). NMR signal reception: Virtual photons and coherent spontaneous emission. Concepts in Magnetic Resonance, 9(5), 277–297.
    [8] Keeler, J. (2011). Understanding NMR Spectroscopy. Wiley.
    [9] Milford, D., Rosbach, N., Bendszus, M., & Heiland, S. (2015). Mono-exponential fitting in T2-relaxometry: relevance of offset and first echo. PLOS One, 10(12).
    [10] Hahn, E. L. (1950). Spin echoes. Phys. Rev. 80(4), 580–594.
    [11] Gujar, S. K., Maheshwari, S., Björkman-Burtscher, I., & Sundgren, P. C. (2005). Magnetic resonance spectroscopy. Journal of Neuro-Ophthalmology, 25(3), 217–226.
    [12] Hore, P. J. (2015). Nuclear Magnetic Resonance. Oxford University Press.
    [13] Hupfeld, K. E., Murali-Manohar, S., Zollner, H. J., Song, Y., Davies-Jenkins, C. W., Gudmundson, A. T., Simicic, D., Simegn, G., Carter, E. E., Hui, S. C. N., Yedavalli, V., Oeltzschner, G., Porges, E. C., & Edden, R. A. E. (2025). Metabolite T2 relaxation times decrease across the adult lifespan in a large multi-site cohort. Magnetic Resonance in Medicine, 93(3), 916–929.
    描述: 碩士
    國立政治大學
    應用物理研究所
    111755003
    資料來源: http://thesis.lib.nccu.edu.tw/record/#G0111755003
    資料類型: thesis
    顯示於類別:[應用物理研究所 ] 學位論文

    文件中的檔案:

    檔案 大小格式瀏覽次數
    500301.pdf1938KbAdobe PDF0檢視/開啟


    在政大典藏中所有的資料項目都受到原著作權保護.


    社群 sharing

    著作權政策宣告 Copyright Announcement
    1.本網站之數位內容為國立政治大學所收錄之機構典藏,無償提供學術研究與公眾教育等公益性使用,惟仍請適度,合理使用本網站之內容,以尊重著作權人之權益。商業上之利用,則請先取得著作權人之授權。
    The digital content of this website is part of National Chengchi University Institutional Repository. It provides free access to academic research and public education for non-commercial use. Please utilize it in a proper and reasonable manner and respect the rights of copyright owners. For commercial use, please obtain authorization from the copyright owner in advance.

    2.本網站之製作,已盡力防止侵害著作權人之權益,如仍發現本網站之數位內容有侵害著作權人權益情事者,請權利人通知本網站維護人員(nccur@nccu.edu.tw),維護人員將立即採取移除該數位著作等補救措施。
    NCCU Institutional Repository is made to protect the interests of copyright owners. If you believe that any material on the website infringes copyright, please contact our staff(nccur@nccu.edu.tw). We will remove the work from the repository and investigate your claim.
    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回饋